FEATURES & SPECIFICATIONS
DESCRIPTION:
The IG-28DS Touch spectrophotometer measures the amount of light that a sample absorbs by passing a light beam through it. These instruments are used to measure the concentration of any unknown solution and monitor absorption accuracy throughout production.
KEY FEATURES:
-
-
- Wide Wavelength Range: Covers a broad spectrum 190-1100 nm, meeting the needs of diverse industries.
- Fully Automated Design: Simplifies measurements and ensures compliance with pharmacopoeia standards.
- Upgradable Sample Capacity: Supports measurement of up to 8 samples simultaneously.
- Automatic Lamp Switching: Seamless transition between Tungsten (T) lamp and Deuterium (D2) lamp for
convenience - Advanced Optical System:
Optimized optics and integrated circuits for precision.
Light sources and detectors ensuring reliability. - Versatile Measurement Modes:
Wavelength Scan
Time Scan
Multi-Wavelength Determination
Multi-Order Derivative Determination
Double Wavelength Method
Triple Wavelength Method - Data Output Options:
Printer port
USB interface
Data storage for easy access and record keeping. - PC Control Compatibility: Enables more accurate and flexible measurement operations through software
integration.
MEASUREMENT
Basic Mode
To measure the Absorbance and transmittance
Quantitative Coefficient Method
Standard Curve Up to 10 Standard sample may be used to establish a Curve. Four methods for fitting a curve through the calibration points: Linear fit, Linear fit through zero, Square fit and cubic fit.
DNA/Protein Test
- Concentration and DNA purity are quickly and easily calculated:
Absorbance ratios 260 nm / 280 nm with optional subtracted absorbance at 320 nm. - DNA concentration = 62.9 × A260 − 36.0 × A280
- Protein concentration = 1552 × A260 − 757.3 × A280
Wavelength Scan
- The wavelength scan intervals are 0.1, 0.2, 0.5, 1, 2, 5 nm.
- High, medium and low scan speeds are available: 100 to 3600 nm/min.
- Wavelengths are scanned from high to low so that the instrument
waits at high wavelength and minimizes the degradation of UV-sensitive samples.
STANDARD ACCESSORIES:- Glass Cell: 4 Nos
- Quartz Cell: 2 Nos
- Software
- Instrument Cover
- USB Cable
- Software Manual
- Instrument Manual
TECHNICAL SPECIFICATION
MODEL IG-28DS Touch Optical System Double Beam, Grating 1200 lines/mm Wavelength Range 190 nm – 1100 nm Spectral Bandwidth 0.5/1/2/4/5 nm Wavelength Accuracy ≤±0.1 nm (656.1 nm D2), ≤±0.3 nm (full wavelength range) Wavelength Repeatability 0.1 nm Photometric Accuracy ±0.3%T (0~100%T) Photometric Repeatability 0.001 Abs (0~0.5 Abs) Photometric Range -3A~3A Stray Light ≤0.02%T (220 nm, natural air level (NaI), 340 nm NaMO2 Sodium Metal Oxide) Stability ±0.0004 A/H @ 500 nm Baseline Flatness ±0.001 A Noise 0.0003 A/H Scanning Speed Fast, Mid, Slow Wavelength Setting Auto Control Panel Touchscreen interface Light Source Deuterium & Tungsten Lamp Wavelength Resolution 0.1 nm Photometric Mode Absorbance, Transmittance and Concentration (A, T and C) Detector Imported Silicon Photodiode Interface USB Port and parallel port (Printer) Power AC 220V/50Hz or AC 110V/60Hz Dimension 590 × 460 × 220 mm Weight 25 Kg Drift ±0.0004 Abs/h Cell Holder 2 Cell Holder -